Use Able AI chat for product recommendations

Speedy™ Human MPO One-Step ELISA Kit

Sensitivity

7.2 pg/mL

Range

39-2500 pg/mL

Reactivity

Human

Cat no : SE50024

Synonyms

EC:1.11.2.2, myeloperoxidase



Product Information

Product name Speedy™ Human MPO One-Step ELISA Kit
Tests 1 X 96 well plate
Sample type Serum, Plasma, Cell culture supernatant, Urine, Saliva, Cell lysate
Assay type Sandwich
Sensitivity 7.2 pg/mL
Range 39-2500 pg/mL
Reactivity Human
Tested applications Sandwich ELISA
Gene ID (NCBI) 4353

Recovery

Sample TypeAverageRange
Human serum108%101%-121%
Cell culture supernatant105%91%-118%
Urine109%108%-110%
Saliva90%88%-95%
Cell lysate110%100%-127%

IntraAssay

Samplenmean ( pg/mL)SDCV%
181,242.7 31.5 2.5
28292.6 7.5 2.6
38143.4 5.1 3.5

InterAssay

Samplenmean ( pg/mL)SDCV%
1161,259.0 34.6 2.8
216307.2 16.8 5.5
316149.2 8.5 5.7

Background Information

MPO(myeloperoxidase) is a peroxidase enzyme presented in the azurophilic granules of polymorphonuclear (PMN) leukocytes and monocytes. Plasma concentration of MPO can be used as a specific marker of PMN activation. MPO catalyzes the production of hypochlorous acid (HClO) from hydrogen peroxide (H2O2) and chloride anion (Cl-, or the equivalent from a non-chlorine halide). This enzymatic system plays an important role in human defense against microorganisms. The serum/plasma MPO levels have been associated with a variety of clinical conditions including inflammatory diseases, atherosclerosis, ischaemic stroke, hypertension, heart failure, risk of cardiovascular events and so on.

Properties

Storage Instructions All the reagents are stored at 2-8℃ for 6 months or -20℃ for 12 months. Refer to the protocol for further storage instructions.
Synonyms EC:1.11.2.2, myeloperoxidase

FAQ

1、I have limited sample diluent. How should I handle samples requiring high dilution factors?

Answer: We recommend performing a serial dilution. For example, if human serum requires a 1,000,000-fold dilution, you can use a three-step 100× dilution method as follows:
Step 1: Take 2 uL of human serum and add it to 198 uL of sample diluent, mix thoroughly to obtain a 100-fold diluted sample.
Step 2: Take 2 uL of the 100-fold diluted sample and add it to 198 uL of sample diluent, mix thoroughly to obtain a 10,000-fold diluted sample.
Step 3: Take 2 uL of the 10,000-fold diluted sample and add it to 198 uL of sample diluent, mix thoroughly to obtain a 1,000,000-fold diluted sample.
If further dilution is required, additional serial steps can be used. We do not recommend using any other solutions or water for dilution, except for the sample diluent we suggest. If your sample diluent is still insufficient, please contact our customer service for more sample diluent.

2、If thermostatic shaker is not available, will it affect the duration of my experiment and the experimental data?

Answer: If thermostatic shaker is not available, it may take 2.5 hours to complete this experiment. After adding the antibody cocktail, we suggest you incubate the plate at 37℃ for 2 hours. After adding TMB, you only need to incubate the plate at 37℃ for 15 minutes. Through comparative experiments, our R&D personnel have confirmed that you can still obtain valid experimental data, although the signal will be slightly reduced.

||